Cylinder filling ULTRAOX + ULTRAFIL

A PSA oxygen plant with cylinder filling.

A PSA oxygen cylinder filling plant makes medical oxygen on site and puts it into cylinders: an ULTRAOX plant produces the oxygen, and an ULTRAFIL filling station compresses it into medical oxygen cylinders at a maximum pressure of 150 to 207 bar, depending on the model.

ULTRAFIL oxygen filling station with ULTRACECO HP booster and two cylinder racks
Filling rack
5 to 20cylinders
Racks of 5, 10 or 20 cylinders
Outlet pressure
207bar
Highest, on ULTRACECO HP3S
Booster capacity
3.2 to 50Nm³/h
Per booster, ULTRACECO HP or HPT
Booster models
11
With double and triple sets
Oxygen purity
93%± 3%
Oxygen 93 grade
How it works

From the plant to the cylinder.

Cylinder filling adds two components to an ULTRAOX plant: a high-pressure oxygen booster and a cylinder filling rack.

01

The plant makes the oxygen.

Compressed air enters one of two zeolite columns. Under pressure, the molecular sieve captures nitrogen molecules while oxygen flows through to the buffer tank, at 93% ± 3% purity.

02

The booster raises the pressure.

An oil-free ULTRACECO HP or ULTRACECO HPT booster takes the oxygen in at 2.5 to 4 bar and compresses it to a maximum of 150 to 207 bar, depending on the model.

03

The rack fills the cylinders.

Filling racks for 5, 10 or 20 cylinders, including double racks of 2 × 5 and 2 × 10, with a stainless flexible hose per line and a rigid frame with safety chains.

The filling station

The ULTRAFIL filling station.

An ULTRAFIL filling station connects to an ULTRAOX plant and fills medical oxygen cylinders using an ULTRACECO HP or ULTRACECO HPT high-pressure oxygen booster, at a maximum pressure of 150 to 207 bar depending on the model. The station has two components: the booster and a cylinder filling rack.

ULTRAOX runs fully autonomously, even when integrated with an ULTRAFIL filling station. If any production parameter deviates from your pre-established thresholds, the system automatically switches to a secondary oxygen source.

ULTRAFIL
Specification
Booster
ULTRACECO HP or ULTRACECO HPT high-pressure oxygen booster
Compression
Oil-free, for medical oxygen
Capacity per booster
3.2 to 50 Nm³/h
Sets
Single, D (two boosters) or T (three boosters)
Outlet pressure
150 to 207 bar, by model
Inlet pressure
2.5 to 4 bar
Filling rack
5, 10 or 20 cylinders, single or double
Power supply
400 V / 50 Hz / 3 AC
Country of origin
PT (EU)
Choosing the booster

ULTRACECO HP or HPT.

ULTRACECO HP covers 3.2 to 17.2 Nm³/h per booster at 150 to 207 bar. ULTRACECO HPT covers 5 to 50 Nm³/h per booster at 150 bar.

BoosterULTRACECO HPULTRACECO HPT
ModelsHP2, HP3 and HP3S. Sets: HP3D, HP3SD, HP3T and HP3STHPT1 to HPT8. Sets: HPT1D to HPT8D and HPT1T to HPT8T
Capacity per booster3.2 Nm³/h on HP2, 17.2 Nm³/h on HP3 and HP3S5 to 50 Nm³/h
Outlet pressure150 bar on HP2, 172 bar on HP3, 207 bar on HP3S150 bar
Inlet pressure2.5 to 4 bar2.5 to 4 bar
Motor power per booster1.1 kW on HP2, 7.5 kW on HP3 and HP3S3 to 18.5 kW
CompressionOil-free, for medical oxygenOil-free, for medical oxygen
The cylinders

Medical oxygen cylinders.

High pressure oxygen cylinders from 2 to 50 L, at 200 or 300 bar working pressure.

High pressure gas cylinders for medical oxygen, in 34CrMo4 steel, made to ISO 9809-2:2010. Cylinders are manufactured in Europe with a TPED certificate, internally surface treated and cleaned to receive direct medical oxygen. Cylinder frames can be sized according to the needs of each customer.

Oxygen per cylinder
Approximate
2 L, 200 bar
427 L
10 L, 200 bar
2,132 L
20 L, 200 bar
4,264 L
40 L, 200 bar
8,532 L
50 L, 200 bar
10,665 L
50 L, 300 bar
15,295 L
Sizing

Sized on cylinders a day.

A plant that fills cylinders is sized on the cylinders: their size and how many must be filled each day.

The ULTRAOX SC and ST datasheets give each plant model a maximum daily filling capacity, counted in 50 L cylinders filled at 150 bar: from 10 cylinders a day on ULTRAOX 30 to 528 on ULTRAOX 1640.

Plants are modular: one or more generators and compressors are combined to match the hospital’s demand, and capacity can be added later without replacing what is already installed. To size a plant, Ultra Controlo needs the oxygen flow required in m³/h or, for cylinder filling, the size and number of cylinders to fill per day.

Max. daily filling capacity
50 L @ 150 bar
ULTRAOX 30
10
ULTRAOX 100
30
ULTRAOX 260
84
ULTRAOX 560
178
ULTRAOX 1120
356
ULTRAOX 1640
528

Indicative, from the datasheet. Our engineers confirm the configuration for your site.

Why ULTRAFIL

Built for medical oxygen.

Oil-free compression

For medical oxygen, with automatic start and stop from the inlet pressure.

A protected fill cycle

High temperature and overpressure shutdowns on the booster, and high pressure relief valve protection on the rack.

New or existing plants

Preassembled manifolds, pressure-rated valves and stainless-steel flexibles enable quick commissioning and clean tie-in to new or existing plants.

Real-world deployments

Filling cylinders on site.

ULTRAOX plants with cylinder filling, from regional hospitals to oxygen stations that supply the health units around them.

TL
February 2026
Timor-Leste · Manatuto

ULTRAOX plant opens in Manatuto, with The Global Fund

The oxygen production container facility has the capacity to store up to 1,500 oxygen cylinders and fill approximately 50 medical oxygen cylinders per day using an ULTRAOX medical oxygen PSA generator.

The Global Fund Read the story
CD
September 2025
DR Congo · Kenge and Lisala

Oxygen production and cylinder-filling systems for Kenge and Lisala

Two ULTRAOX systems arrived at Port TFCE in Kinshasa for the hospitals of Kenge and Lisala, as part of the national Lot 2 project for three complete oxygen production and cylinder-filling systems.

CD
November 2023
DR Congo · Mbujimayi

ULTRAOX plant inaugurated at BONZOLA General Reference Hospital

The ULTRAOX plant inaugurated on 3 November 2023 at the BONZOLA General Reference Hospital in Mbujimayi, capital of Kasai Oriental, can fill up to 50 cylinders of 50 L every day.

50 cylinders a day Read the story
PE
June 2021
Peru · San Martín

Two ULTRAOX plants able to fill 95 cylinders a day

The Regional Government of San Martín acquired two ULTRAOX medical oxygen generator systems, able to produce 40 m³ of oxygen per hour and to fill 95 cylinders of 10 m³ each per day, for the hospitals of Tarapoto and Moyobamba.

95 cylinders a day Read the story
PE
June 2021
Peru · Tingo María

Two filling lines at Tingo María Hospital

The ULTRAOX plant at Tingo María Hospital has two filling lines for 100 oxygen cylinders a day, to help meet the demand of the population of Leoncio Prado.

Two filling lines Read the story
AO
February 2021
Angola · Sumbe

An oxygen station for the health units of Cuanza-Sul

Hospital 17 de Setembro in Sumbe, opened by the Minister of Health, has an ULTRAOX oxygen station that can fill 50 cylinders a day and supply all local healthcare units in and around Cuanza-Sul Province.

50 cylinders a day Read the story
Frequently asked questions

Questions about cylinder filling.

Questions hospitals ask us about on-site oxygen. More on the medical gas answers page.

Can a PSA oxygen plant fill medical oxygen cylinders?

Yes: an ULTRAFIL filling station connects to an ULTRAOX plant and fills medical oxygen cylinders using an ULTRACECO HP or ULTRACECO HPT high-pressure oxygen booster, at a maximum pressure of 150 to 207 bar depending on the model. The filling rack takes 5, 10 or 20 cylinders, in single or double racks.

What does an ULTRAFIL station include?

A high-pressure oxygen booster, ULTRACECO HP or ULTRACECO HPT, and a cylinder filling rack.

How many cylinders a day can an ULTRAOX plant fill?

The ULTRAOX SC and ST datasheets give each plant model a maximum daily filling capacity, counted in 50 L cylinders filled at 150 bar: from 10 cylinders a day on ULTRAOX 30 to 528 on ULTRAOX 1640.

For oxygen, give the flow in m³/h or, for cylinder filling, the cylinder size and how many cylinders a day. Ultra Controlo’s engineering team sizes the system from those figures.

Should I choose HP or HPT?

ULTRACECO HP covers 3.2 to 17.2 Nm³/h per booster at 150 to 207 bar. ULTRACECO HPT covers 5 to 50 Nm³/h per booster at 150 bar.

Every size from HPT1 to HPT8 comes as a single booster, a D set of two or a T set of three.

HP3 and HP3S come in D sets of two boosters and T sets of three. HP2 is single only.

Can ULTRAFIL be added to a plant that is already installed?

The filling rack is designed for clean tie-in to new or existing plants. Booster inlet pressure is 2.5 to 4 bar.

Which cylinders does Ultra Controlo supply?

High pressure oxygen cylinders from 2 to 50 L, at 200 or 300 bar working pressure.

High pressure gas cylinders for medical oxygen, in 34CrMo4 steel, made to ISO 9809-2:2010. Cylinders are manufactured in Europe with a TPED certificate, internally surface treated and cleaned to receive direct medical oxygen. Cylinder frames can be sized according to the needs of each customer.

Request a project assessment

Ready to discuss oxygen for your hospital?

Tell us about your site. Our engineering team sizes the configuration and sends a quote.

ISO 9001:2015
ISO 14001
ISO 13485:2016
80+ countries
40 years