ULTRAAR SCA is a modular, oil-free medical air production system that meets the specifications of ISO 7396-1 (medical gas pipeline systems). Oil-exempt scroll compressor groups feed an ULTRASEC refrigeration dryer and a duplex pressure reducing kit with sterilisation filter.
Medical CE marked (CE 1639) under Directive 93/42/EEC. Meets the specifications of ISO 7396-1, manufactured under ISO 13485:2016.
ULTRACECO SCA groups with high efficiency motors, class F insulation, V-belt drive and a cyclonic separator with electronic condensate drain.
All the main components are prepared in the factory, and every system is tested and predefined before it leaves.
Electrical control panel with a Manual, Off and Automatic selector, run and failure signals and an hour meter per compressor, plus potential-free alarms for remote signalling.
| Model | bar | psig | l/s | m³/h | cfm | kW | L × W × H mm | kg |
|---|---|---|---|---|---|---|---|---|
| SCA 1 | 8 | 116 | 2.7 | 9.6 | 5.7 | 1.5 | 650 × 800 × 630 | 55 |
| SCA 2 | 8 | 116 | 4 | 14.4 | 8.5 | 2.2 | 650 × 800 × 630 | 60 |
| SCA 2 | 10 | 145 | 3.4 | 12 | 7.1 | 2.2 | 650 × 800 × 630 | 60 |
| SCA 4 | 8 | 116 | 6.7 | 24 | 14.1 | 4 | 1,010 × 720 × 740 | 92 |
| SCA 5 | 8 | 116 | 10 | 36 | 21.2 | 5.5 | 1,010 × 720 × 740 | 93 |
| SCA 5 | 10 | 145 | 7.4 | 26.4 | 15.6 | 5.5 | 1,010 × 720 × 740 | 93 |
| SCA 7 | 8 | 116 | 14.2 | 51 | 30 | 7.5 | 1,010 × 720 × 740 | 93 |
| SCA 21 | 8 | 116 | 5.4 | 19.2 | 11.4 | 3 | 650 × 800 × 1,120 | 120 |
| SCA 22 | 8 | 116 | 8 | 28.8 | 17 | 4.4 | 650 × 800 × 1,120 | 130 |
| SCA 22 | 10 | 145 | 6.8 | 24 | 14.2 | 4.4 | 650 × 800 × 1,120 | 130 |
| SCA 24 | 8 | 116 | 13.4 | 48 | 28.2 | 8 | 1,010 × 720 × 1,340 | 188 |
| SCA 25 | 8 | 116 | 20 | 72 | 42.4 | 11 | 1,010 × 720 × 1,340 | 190 |
| SCA 25 | 10 | 145 | 14.8 | 52.8 | 31.2 | 11 | 1,010 × 720 × 1,340 | 190 |
| SCA 27 | 8 | 116 | 28.4 | 102 | 60 | 15 | 1,010 × 720 × 1,340 | 190 |
| SCA 31 | 8 | 116 | 8.1 | 28.8 | 17.1 | 4.5 | 650 × 800 × 1,610 | 178 |
| SCA 32 | 8 | 116 | 12 | 43.2 | 25.5 | 6.6 | 650 × 800 × 1,610 | 193 |
| SCA 32 | 10 | 145 | 10.2 | 36 | 21.3 | 6.6 | 650 × 800 × 1,610 | 193 |
| SCA 34 | 8 | 116 | 20.1 | 72 | 42.3 | 12 | 1,010 × 720 × 1,940 | 284 |
| SCA 35 | 8 | 116 | 30 | 108 | 63.6 | 16.5 | 1,010 × 720 × 1,940 | 287 |
| SCA 35 | 10 | 145 | 22.2 | 79.2 | 46.8 | 16.5 | 1,010 × 720 × 1,940 | 287 |
| SCA 37 | 8 | 116 | 42.6 | 153 | 90 | 22.5 | 1,010 × 720 × 1,940 | 287 |
| SCA 41 | 8 | 116 | 10.8 | 38.4 | 22.8 | 6 | 600 × 1,600 × 1,120 | 233 |
| SCA 42 | 8 | 116 | 16 | 57.6 | 34 | 8.8 | 600 × 1,600 × 1,120 | 253 |
| SCA 42 | 10 | 145 | 13.6 | 48 | 28.4 | 8.8 | 600 × 1,600 × 1,120 | 253 |
| SCA 44 | 8 | 116 | 26.8 | 96 | 56.4 | 16 | 1,110 × 1,440 × 1,340 | 344 |
| SCA 45 | 8 | 116 | 40 | 144 | 84.8 | 22 | 1,110 × 1,440 × 1,340 | 380 |
| SCA 45 | 10 | 145 | 29.6 | 105.6 | 62.4 | 22 | 1,110 × 1,440 × 1,340 | 380 |
| SCA 47 | 8 | 116 | 56.8 | 204 | 120 | 30 | 1,110 × 1,440 × 1,340 | 380 |
Capacities measured according to ISO 1217 (displacement compressors: acceptance tests), Annex C. Reference conditions: absolute inlet pressure 1 bar (14.5 psig), suction temperature 20 °C. Motor 50/60 Hz, 208 to 440 V.
4 photos of the ULTRAAR SCA




Yes. It is a modular, oil-free medical air production system with oil-exempt scroll compressor groups.
ULTRACECO SCA compressors, an ULTRASEC refrigeration dryer, an electrical control panel and a duplex pressure reducing kit with sterilisation filter.
In automatic mode the compressors alternate to even out wear. Each compressor has its own pressure switch for automatic start and stop at the preset pressure.



Ultra Controlo is formed by a dynamic and enterprising team dedicated to the manufacture of medical gas plants, assembled and tested in our factories and ready to be installed in clinics and hospitals everywhere around the world.
We supply complete turn-key projects, build to fit your needs, on Medical Gas Piping Systems (MGPS).