Sizes 3.10 to 165.3 Nm³/h

PSA oxygen generator sizes, ULTRAOX 30 to 1640.

ULTRAOX PSA oxygen generators come in 16 sizes, from 3.10 Nm³/h of oxygen on ULTRAOX 30 to 165.3 Nm³/h on ULTRAOX 1640. Find your flow in the table, and Ultra Controlo engineers confirm the configuration for your site.

ULTRAOX medical oxygen plant: air and oxygen tanks, ULTRACECO compressor, air treatment, ULTRACON generators and the backup cylinder bank
Models
16sizes
ULTRAOX 30 to 1640
Oxygen flow
3.10to 165.3
Nm³/h of oxygen
Oxygen purity
93%± 3%
Oxygen 93 grade
Altitude
3bands
Compressor sized up to 3,000 m
Cylinders a day
528max.
ULTRAOX 1640, 50 L @ 150 bar
The 16 sizes

Every ULTRAOX size, by oxygen flow.

Each model is listed with its oxygen flow in Nm³/h, the normal cubic metres per hour. ULTRAOX SC and ULTRAOX ST have the same 16 models, oxygen flows and ULTRACON generators.

Simplex layout. Same oxygen capacity in all three altitude bands. Indicative, from the datasheet. Our engineers confirm the configuration for your site. ULTRAOX 520, 730, 1120 and 1640 have two ULTRACON generators of the same model, written × 2.

Normal cubic metre (Nm³)

ModelOxygen, Nm³/hMax. daily filling capacity, cylinders (50 L @ 150 bar)Oxygen generator
ULTRAOX 303.1010ULTRACON 4
ULTRAOX 504.6014ULTRACON 5
ULTRAOX 706.5020ULTRACON 6
ULTRAOX 908.0025ULTRACON 7
ULTRAOX 1009.5030ULTRACON 8
ULTRAOX 12012.440ULTRACON 9
ULTRAOX 18018.258ULTRACON 10
ULTRAOX 26026.284ULTRACON 11
ULTRAOX 37036.5116ULTRACON 12
ULTRAOX 44043.5139ULTRACON 13
ULTRAOX 52052.4168ULTRACON 11 × 2
ULTRAOX 56055.7178ULTRACON 14
ULTRAOX 73073.0232ULTRACON 12 × 2
ULTRAOX 82082.7264ULTRACON 15
ULTRAOX 1120111.4356ULTRACON 14 × 2
ULTRAOX 1640165.3528ULTRACON 15 × 2
Compressor by altitude

The ULTRAOX ST compressor, by altitude band.

Each model has its compressor size for three altitude bands: 0 to 1,000 m, 1,000 to 2,000 m and 2,000 to 3,000 m. The compressor is an ULTRASCREW VSD.

The oxygen flow of a model is the same in all three bands. At 2,000 to 3,000 m, ULTRAOX 1640 uses 2 × ULTRASCREW 160, with 2 oxygen filtration systems and 2 flow measurement systems.

Derating for altitude and site conditions

Model0 to 1,000 m1,000 to 2,000 m2,000 to 3,000 m
ULTRAOX 30ULTRASCREW 5ULTRASCREW 7ULTRASCREW 7
ULTRAOX 50ULTRASCREW 7ULTRASCREW 11ULTRASCREW 11
ULTRAOX 70ULTRASCREW 11ULTRASCREW 15ULTRASCREW 15
ULTRAOX 90ULTRASCREW 15ULTRASCREW 18ULTRASCREW 18
ULTRAOX 100ULTRASCREW 18ULTRASCREW 18ULTRASCREW 18
ULTRAOX 120ULTRASCREW 18ULTRASCREW 22ULTRASCREW 30
ULTRAOX 180ULTRASCREW 30ULTRASCREW 37ULTRASCREW 37
ULTRAOX 260ULTRASCREW 37ULTRASCREW 45ULTRASCREW 55
ULTRAOX 370ULTRASCREW 55ULTRASCREW 75ULTRASCREW 75
ULTRAOX 440ULTRASCREW 75ULTRASCREW 75ULTRASCREW 75
ULTRAOX 520ULTRASCREW 75ULTRASCREW 90ULTRASCREW 110
ULTRAOX 560ULTRASCREW 75ULTRASCREW 110ULTRASCREW 110
ULTRAOX 730ULTRASCREW 110ULTRASCREW 132ULTRASCREW 160
ULTRAOX 820ULTRASCREW 132ULTRASCREW 160ULTRASCREW 160
ULTRAOX 1120ULTRASCREW 160ULTRASCREW 200ULTRASCREW 200
ULTRAOX 1640ULTRASCREW 250ULTRASCREW 3152 × ULTRASCREW 160
SC, ST and CO

Same sizes, two compressors.

ULTRAOX SC, ST and CO share the same PSA oxygen generators and differ in how the air is compressed and where the plant is housed.

ULTRAOX SC uses oil-less ULTRACECO compressors. ULTRAOX ST uses ULTRASCREW rotary screw compressors. ULTRAOX CO is a complete plant built inside a shipping container, for sites with no technical room or no space for civil works.

Containerised PSA oxygen plant, ULTRAOX CO

ULTRAOX SC, ST and CO
Compared
ULTRAOX SC
ULTRACECO oil-less scroll, Class 0
ULTRAOX ST
ULTRASCREW oil-injected rotary screw
ULTRAOX CO
Complete plant inside a shipping container
Oxygen flow per unit
ULTRAOX SC and ST
3.10 to 165.3 Nm³/h
ULTRAOX CO
Sized on request
Getting a size confirmed

What to send for a quote.

To size a plant, Ultra Controlo needs the oxygen flow required in m³/h or, for cylinder filling, the size and number of cylinders to fill per day.

01

Your oxygen flow

Send the oxygen flow you need in m³/h, or use the table above to find the closest size.

02

Cylinders per day

If the plant will also fill cylinders, give the cylinder size and how many cylinders a day.

03

Your site

Heat, humidity, dust, altitude and the power supply all change what a PSA plant needs, so give the manufacturer the real site conditions.

Frequently asked questions

Questions buyers ask about sizes.

Questions hospitals ask us about on-site oxygen. More on the medical gas answers page.

What sizes of PSA oxygen generator does Ultra Controlo make?

ULTRAOX plant models range from 3.10 Nm³/h of oxygen on ULTRAOX 30 to 165.3 Nm³/h on ULTRAOX 1640, delivered at 5 to 6 bar; each plant has one or two ULTRACON generators.

There are 16 models, listed in the table above with their oxygen flow, daily cylinder filling and ULTRACON generator.

Which ULTRAOX size gives 30 or 60 Nm³/h?

The datasheet table has no model at exactly 30 or 60 Nm³/h. It lists ULTRAOX 260 at 26.2 Nm³/h and ULTRAOX 370 at 36.5 Nm³/h, and ULTRAOX 560 at 55.7 Nm³/h and ULTRAOX 730 at 73.0 Nm³/h.

Indicative, from the datasheet. Our engineers confirm the configuration for your site.

What flow is the ULTRAOX 1640?

ULTRAOX 1640 is listed at 165.3 Nm³/h of oxygen, with two ULTRACON 15 generators and a maximum daily filling capacity of 528 cylinders (50 L @ 150 bar).

Does altitude change the size of the plant?

Each model has its compressor size for three altitude bands: 0 to 1,000 m, 1,000 to 2,000 m and 2,000 to 3,000 m.

Same oxygen capacity in all three altitude bands.

How many cylinders can an ULTRAOX plant fill a day?

The tables list a maximum daily filling capacity from 10 cylinders on ULTRAOX 30 to 528 cylinders on ULTRAOX 1640, for cylinders of 50 L @ 150 bar. See the PSA oxygen cylinder filling plant page.

Can capacity be added later?

Plants are modular: one or more generators and compressors are combined to match the hospital’s demand, and capacity can be added later without replacing what is already installed.

How do I get a size and a quote?

Email leads@ultra-controlo.com or call +351 219 154 350 with the gases you need, the site and the demand.

For oxygen, give the flow in m³/h or, for cylinder filling, the cylinder size and how many cylinders a day. Ultra Controlo’s engineering team sizes the system from those figures.

Request a project assessment

Ready to discuss oxygen for your hospital?

Tell us about your site. Our engineering team sizes the configuration and sends a quote.

CE 1639
Class IIb
ISO 13485:2016
80+ countries
40 years